MicroSet Instrumentation & Controls, Electromagnetic Flow Meter, PH Meters, Slim PH Meter, PH Sensor, Sterilizable PH Sensor, Autoclavable PH Sensor, PH Controller, PH Meter, Conductivity Sensor, Conductivity Controller

22 years Celebrating


Manufacturer, Supplier and Exporter of Fermentation Bioreactor Sensors and our setup is situated in Pune, Maharashtra, India.

Digital TDS & Conductivity Meter MS CD 610622
The dissolved oxygen sensor offers the most rugged design for 25 mm side ports. It fits directly into the vessel, eliminating the need for a housing, reducing contamination risks. The modular design allows quick replacement of any part, making it easy to maintain and repair the sensor.

FEATURE
  • 316L Stainless Steel
  • Sterialized for traceability
  • RA32 Finish with Electropolish on wetted parts
  • Internal and External Silicone O-rings
  • Secured retainer ring
PERFORMANCE SPECIFICATION
Measurement principal
:
Amperometric – Polarographic
Output
:
25 … 50nA
Output signal in air
:
Approx. 67 nA at 25°C
Temperature range
:
Operation: 0 to 80°C
 
:
Sterilization: Max. 135°C
Stability
:
Better than 2% of readout per week at constant pressure and temp (30°C)
Maximum Pressure
:
4 bar (atm), 58 psig
Oper Temperature Range
:
0 to 80°C
Shaft material
:
316SS
Shaft diameter
:
25 mm
Cable/Connector
:
4-pin




pdf-download

Catalogue